Detailed Images of Deep Brain Stimulation Leads Using Micro-CT
Abstract
Introduction:
One of the challenges in directional deep brain stimulation (DBS) is to determine the orientation of implanted electrodes relative to targeted regions. Post-operative images must be aligned with a model of the implanted lead, usually a computer-based model provided by the manufacturer. This paper shows that models can alternatively be obtained by capturing images of individual leads using micro-CT, a high-resolution CT technique. Contrary to computer-aided design models, lead models generated this way provide realistic X-ray contrast and finer details.
Methods:
We scanned DBS leads from various vendors using a Bruker SkyScan 1276 micro-CT system. To reduce beam-hardening artefacts, samples were scanned at maximum X-ray tube voltage (100 kV) and with copper filtering. Images were made publicly available for download and 3D visualisation.
Conclusion:
Detailed images of single DBS leads can be generated using standard micro-CT systems. Their use as reference models could improve lead orientation algorithms, in particular those dedicated to X-ray modalities. Furthermore, the possibility to share models online could broaden access for clinical research.
One of the challenges in directional deep brain stimulation (DBS) is to determine the orientation of implanted electrodes relative to targeted regions. Post-operative images must be aligned with a model of the implanted lead, usually a computer-based model provided by the manufacturer. This paper shows that models can alternatively be obtained by capturing images of individual leads using micro-CT, a high-resolution CT technique. Contrary to computer-aided design models, lead models generated this way provide realistic X-ray contrast and finer details.
Methods:
We scanned DBS leads from various vendors using a Bruker SkyScan 1276 micro-CT system. To reduce beam-hardening artefacts, samples were scanned at maximum X-ray tube voltage (100 kV) and with copper filtering. Images were made publicly available for download and 3D visualisation.
Conclusion:
Detailed images of single DBS leads can be generated using standard micro-CT systems. Their use as reference models could improve lead orientation algorithms, in particular those dedicated to X-ray modalities. Furthermore, the possibility to share models online could broaden access for clinical research.