Abstract
We demonstrate the development of a label-free, impedance-based biosensor by using a passivation layer of 50-nm tantalum pentoxide (Ta2O5) on interdigitated electrodes (IDE). This layer was fabricated by atomic layer deposition (ALD) and has a high dielectric constant (high-κ), which improves the capacitive property of the IDE. We validate the biosensor’s performance by measuring uromodulin, a urine biomarker for kidney tubular damage, from artificial urine samples. The passivation layer is functionalized with uromodulin antibodies for selective binding. The passivated IDE enables the non-faradaic impedance measurement of uromodulin concentrations with a measurement range from 0.5 ng/mL to 8 ng/mL and with a relative change in impedance of 15 % per ng/mL at a frequency of 150 Hz (log scale). This work presents a concept for point-of-care biosensing applications for disease biomarkers.
Author(s)
Vora, Kunj Himanshu  
Fraunhofer-Institut für Mikroelektronische Schaltungen und Systeme IMS  
Kordas, Norbert  
Fraunhofer-Institut für Mikroelektronische Schaltungen und Systeme IMS  
Seidl, Karsten  
Fraunhofer-Institut für Mikroelektronische Schaltungen und Systeme IMS  
Journal
Sensors. Online journal  
Project(s)
Urinschnelltest für die Erkennung von chronischen Nierenerkrankungen
Funder
Bundesministerium für Bildung und Forschung  
Open Access
File(s)
Download (775.05 KB)
Rights
CC BY 4.0: Creative Commons Attribution
DOI
10.3390/s23249696
10.24406/publica-2310
Additional link
Full text
Language
English
Fraunhofer-Institut für Mikroelektronische Schaltungen und Systeme IMS  
Keyword(s)
  • impedance spectroscopy

  • atomic layer deposition

  • tantalum pentoxide

  • uromodulin

  • chronic kidney disease