2000
Journal Article
Title

3D heart modelling from biplane, rotational angiocardiographic X-ray sequences

Abstract
The 3DHeartView project produced an advanced software application for time-variant 3D cardiac modelling based on back projection of rotational X-ray angiographic sequences. Due to standard DICOM 3.0 interface, the software can be used as an add-on to any modern digital angiographer. The system accuracy has been measured and the approach has been validated with static and dynamic phantom object studies. The prototype has been tested in a clinical routine environment as well. In this paper we present the main features of the system, the methodology of 3D angiographic modelling, and certain testing and operation results. We also describe the application of high-performance computing technology in the modelling process.
Author(s)
Kehl, H.G.
Jäger, J.
Gehrmann, J.
Vogt, J.
Sakas, G.
Fraunhofer-Institut für Graphische Datenverarbeitung IGD  
Kassenböhmer, R.
Papazis, N.
Dimitrelos, D.
Journal
Computers and Graphics  
DOI
10.1016/S0097-8493(00)00075-3
Language
English
Fraunhofer-Institut für Graphische Datenverarbeitung IGD  
Keyword(s)
  • back projection

  • 3D modelling

  • cardiology

  • Cookie settings