The image quality of magnetic particle imaging depends on the interference-free measurement of the particle spectrum. Systematic errors due to harmonic distortions in the components used for implementation complicate the development of efficient imaging devices. To test the suitability of parts for the development of MPI scanners, a test system was developed in which the components to be tested can be integrated, to evaluate their suitability using the measured spectrum. Different forms of connectors integrated into the high-current path were analyzed for their influence on the signal quality using this test system. An evaluation of these components can be made from the clearly visible distortions in the spectrum recorded with the high-sensitive test system.