Design study of a Miniaturized Displacement Transducer (MDT) for an active middle ear implant system

Abstract
People suffering from moderate to severe hearing loss can be treated with active middle ear implants. A new approach in this field is to implant an electromechanical transducer onto the round window membrane in order to improve coupling and be able to treat patients with middle-ear problems. In this paper the design study for a miniaturized displacement transducer (MDT) for the round window is presented. Based on a requirement analysis, the basic principle and analytical modeling of the actuator is shown. A parameter variation study results in an optimized actuator configuration that is able to generate an amplification of 110 dB SPL theoretically. As a next step this actuator has to be manufactured and tested.
Author(s)
Kaltenbacher, Dominik
Schächtele, Jonathan
Goll, Erich
UK Tübingen / HNO-Klinik
Burkhardt, Claus
naturam and Medical Sciences Institute, Tübingen
Arnold, Heinz
UK Tübingen / HNO-Klinik
Dalhoff, Ernst
UK Tübingen / HNO-Klinik
Zenner, Hans-Peter
UK Tübingen / HNO-Klinik
Journal
Biomedical microdevices  
Project(s)
ICASImplantat
Funder
Bundesministerium für Bildung und Forschung  
DOI
10.1007/s10544-014-9884-y
Language
English
Fraunhofer-Institut für Produktionstechnik und Automatisierung IPA  
Keyword(s)
  • piezoelectrical actuator

  • Implantat

  • Cookie settings