1996
Conference Paper
Title

Multianalyt sensor for the simultaneous determination of glucose, L-lactate and uric acid based on a microelectronic array

Abstract
In this work a microelectrode array was used to develop a multianalyte sensor. Glucose, L-lactate and uric acid were chosen as model analytes. The array consists of ten groups of 40 microelectrodes connected in parallel, each of a size of 36 mu m. Thus the array showed a large overall current response retaining typical microelectrode features. Single rows of elecrodes were modified with enzymes entrapped in a polymer matrix. As oxidases were used as enzymes hydrogenes peroxide generated by the enzymatic reaction was detected by anodic oxidation at + 800 mV vs Ag/AgCl/3M KCl. The elementation of interferences of a NAFION membrane between the electrode surface and enzyme layer.
Author(s)
Frebel, H.
Chemnitius, G.
Cammann, K.
Kakerow, R.
Rospert, M.
Mokwa, W.
Mainwork
Eurosensors X. Proceedings  
Conference
Eurosensors 1996  
Language
English
Fraunhofer-Institut für Mikroelektronische Schaltungen und Systeme IMS  
Keyword(s)
  • Enzym

  • gas

  • Meßaufnehmer

  • Mikroelektrode

  • Sauerstoff