• English
  • Deutsch
  • Log In
    Password Login
    Research Outputs
    Fundings & Projects
    Researchers
    Institutes
    Statistics
Repository logo
Fraunhofer-Gesellschaft
  1. Home
  2. Fraunhofer-Gesellschaft
  3. Scopus
  4. Influence of a Modified Procedure of Joining Ceramic Head and Adapter Sleeve on the Stem Taper in Revision: An Experimental Study
 
  • Details
  • Full
Options
2024
Journal Article
Title

Influence of a Modified Procedure of Joining Ceramic Head and Adapter Sleeve on the Stem Taper in Revision: An Experimental Study

Abstract
In revision operations, ceramic heads of modular hip implants can be replaced. As the surface of the stem taper can be damaged, additional adapter sleeves are applied. The components are usually connected manually by the surgeon in a one-step procedure by hammer impacts. In this study, we investigated a two-step joining procedure with reproducible impaction force. First, the adapter sleeve and head were joined quasi-statically with a force of 2 kN using an assembly device. In the second step, these components were applied to the stem taper using a pulse-controlled instrument. For reference, the joints were assembled according to standard conditions using a tensile testing machine. An average pull-off force of 1309 ± 201 N was achieved for the components joined by the instrument, and the average measured values for the components joined by the testing machine were 1290 ± 140 N. All specimens achieved a force >350 N when released and therefore met the acceptance criterion defined for this study. This study showed that a modified procedure in two steps with a defined force has a positive effect on the reproducibility of the measured joining forces compared to previous studies.
Author(s)
Hunger, Sandra  
Fraunhofer-Institut für Werkzeugmaschinen und Umformtechnik IWU  
Rotsch, Christian  
Fraunhofer-Institut für Werkzeugmaschinen und Umformtechnik IWU  
Günther, Florian
Fraunhofer-Institut für Werkzeugmaschinen und Umformtechnik IWU  
Drossel, Welf-Guntram  
Fraunhofer-Institut für Werkzeugmaschinen und Umformtechnik IWU  
Heyde, Christoph-Eckhard
Journal
Bioengineering  
Open Access
DOI
10.3390/bioengineering11020170
Language
English
Fraunhofer-Institut für Werkzeugmaschinen und Umformtechnik IWU  
Keyword(s)
  • adapter sleeve

  • ceramic head

  • insertion

  • modular hip implant

  • revision surgery

  • surgical procedure

  • total hip arthroplasty