• English
  • Deutsch
  • Log In
    Password Login
    Research Outputs
    Fundings & Projects
    Researchers
    Institutes
    Statistics
Repository logo
Fraunhofer-Gesellschaft
  1. Home
  2. Fraunhofer-Gesellschaft
  3. Artikel
  4. 144 channel measurement IC for CdZnTe sensors with energy and time resolution
 
  • Details
  • Full
Options
2014
Journal Article
Title

144 channel measurement IC for CdZnTe sensors with energy and time resolution

Abstract
A 144 channel measurement IC for CdZnTe detectors, used for PET, is presented. Each channel consists of a charge sensitive amplifier, a fast and a slow shaper, a peak sampler for the energy acquisition and an event detector based on a time to digital converter to generate an accurate time stamp for each event. The channels are multiplexed to a fast pipeline ADC on demand. Measurement results for the ASIC showed a noise equivalent input charge of 800 e−rms and a time resolution of 737 ps rms. Evaluation results with a CdZnTe detector yielded an energy resolution of 4.4% full width half maximum at 662 keV with a 137Cs radiation source. The IC is implemented on a 180 nm CMOS process with a total chip size of 100 mm2.
Author(s)
Voelker, Matthias
Hauer, Johann
Benlloch, Jose Maria
I3M
Soriano-Asensi, Antonio
I3M, Spain
Sánchez, Filomeno
I3M, Centro Mixto CSIC-UPV-CIEMAT, Spain
Carrascal, Jorge
CIEMAT
Cela, Jose-Manuel
CIEMAT, Madrid
Romero, Luciano
CIEMAT, Madrid
Journal
Microelectronics journal  
Open Access
DOI
10.1016/j.mejo.2014.03.009