Subject-specific timing adaption in time-encoded arterial spin labeling imaging

Abstract
Objectives: One challenge in arterial spin labeling (ASL) is the high variability of arterial transit times (ATT), which causes associated arterial transit delay (ATD) artifacts. In patients with pathological changes, these artifacts occur when post-labeling delay (PLD) and bolus durations are not optimally matched to the subject, resulting in difficult quantification of cerebral blood flow (CBF) and ATT. This is also true for the free lunch approach in Hadamard-encoded pseudocontinuous ASL (H-pCASL).
Material and methods: Five healthy volunteers were scanned with a 3 T MR-system. pCASL-subbolus timing was adjusted individually by the developed adaptive Walsh-ordered pCASL sequence and an automatic feedback algorithm. The quantification results for CBF and ATT and the respective standard deviations were compared with results obtained using recommended timings and intentionally suboptimal timings.
Results: The algorithm individually adjusted the pCASL-subbolus PLD for each subject within the range of recommended timing for healthy subjects, with a mean intra-subject adjustment deviation of 47.15 ms for single-shot and 44.5 ms for segmented acquisition in three repetitions.
Discussion: A first positive assessment of the results was performed on healthy volunteers. The extent to which the results can be transferred to patients and are of benefit must be investigated in follow-up studies.
Author(s)
Breutigam, Nora-Josefin  
Fraunhofer-Institut für Digitale Medizin MEVIS  
Hoinkiss, Daniel Christopher
Fraunhofer-Institut für Digitale Medizin MEVIS  
Konstandin, Simon  
Fraunhofer-Institut für Digitale Medizin MEVIS  
Buck, Mareike
Fraunhofer-Institut für Digitale Medizin MEVIS  
Mahroo, Amnah
Fraunhofer-Institut für Digitale Medizin MEVIS  
Eickel, Klaus  
Fraunhofer-Institut für Digitale Medizin MEVIS  
Samson-Himmelstjerna, Federico von
Fraunhofer-Institut für Digitale Medizin MEVIS  
Günther, Matthias  
Fraunhofer-Institut für Digitale Medizin MEVIS  
Journal
Magnetic resonance materials in physics, biology and medicine : MAGMA  
Project(s)
ERA-NET to support the Joint Programming in Neurodegenerative Diseases strategic plan (JPND)  
Funder
European Commission  
Open Access
DOI
10.1007/s10334-023-01121-y
Additional link
Full text
Language
English
Fraunhofer-Institut für Digitale Medizin MEVIS  
Keyword(s)
  • Arterial spin labeling

  • Arterial transit delay artifacts

  • Free-lunch approach

  • Subject-specific timing

  • Time-encoded pCASL

  • Cookie settings
  • Imprint
  • Privacy policy
  • Api
  • Contact
© 2024