A cw-terahertz gas analysis system with ppm detection limits

Abstract
We present a demonstration system for gas analysis based on cw-terahertz absorption spectroscopy. The setup combines tunable 1.5 mm distributed-feedback (DFB) lasers, InGaAs photomixers and a multipass absorption cell into a high-resolution spectrometer. At gas pressures of 1 mbar, we achieved detection limits in the range of approx. 20 ppm for ammonia (NH3) and 100 ppm for sulfur dioxide (SO2). Absorption profiles with halfwidths of 7 MHz (2×10-4 cm-1) were measured.
Author(s)
Hepp, C.
Lüttjohann, S.
Roggenbuck, A.
Deninger, A.
Nellen, S.
Göbel, T.
Jörger, M.
Harig, R.
Mainwork
41st International Conference on Infrared, Millimeter, and Terahertz waves, IRMMW-THz 2016  
Conference
International Conference on Infrared, Millimeter, and Terahertz Waves (IRMMW-THz) 2016  
DOI
10.1109/IRMMW-THz.2016.7758344