Options
1996
Conference Paper
Titel
Multianalyt sensor for the simultaneous determination of glucose, L-lactate and uric acid based on a microelectronic array
Abstract
In this work a microelectrode array was used to develop a multianalyte sensor. Glucose, L-lactate and uric acid were chosen as model analytes. The array consists of ten groups of 40 microelectrodes connected in parallel, each of a size of 36 mu m. Thus the array showed a large overall current response retaining typical microelectrode features. Single rows of elecrodes were modified with enzymes entrapped in a polymer matrix. As oxidases were used as enzymes hydrogenes peroxide generated by the enzymatic reaction was detected by anodic oxidation at + 800 mV vs Ag/AgCl/3M KCl. The elementation of interferences of a NAFION membrane between the electrode surface and enzyme layer.
Konferenz