A robust algorithm for optic disc segmentation and fovea detection in retinal fundus images

 
: Rust, C.; Häger, S.; Traulsen, N.; Modersitzki, J.

:
Volltext ()

Current directions in biomedical engineering 3 (2017), Nr.2, S.533-537
ISSN: 2364-5504
Englisch
Zeitschriftenaufsatz, Elektronische Publikation
Fraunhofer MEVIS ()

Abstract
Accurate optic disc (OD) segmentation and fovea detection in retinal fundus images are crucial for diagnosis in ophthalmology. We propose a robust and broadly applicable algorithm for automated, robust, reliable and consistent fovea detection based on OD segmentation. The OD segmentation is performed with morphological operations and Fuzzy C Means Clustering combined with iterative thresholding on a foreground segmentation. The fovea detection is based on a vessel segmentation via morphological operations and uses the resulting OD segmentation to determine multiple regions of interest. The fovea is determined from the largest, vessel-free candidate region. We have tested the novel method on a total of 190 images from three publicly available databases DRIONS, Drive and HRF. Compared to results of two human experts for DRIONS database, our OD segmentation yielded a dice coefficient of 0.83. Note that missing ground truth and expert variability is an issue. The new scheme achieved an overall success rate of 99.44% for OD detection and an overall success rate of 96.25% for fovea detection, which is superior to state-of-the-art approaches.