Fraunhofer-Gesellschaft

Publica

Hier finden Sie wissenschaftliche Publikationen aus den Fraunhofer-Instituten.

Intravascular electric impedance spectroscopy of atherosclerotic lesions using a new impedance catheter system

 
: Süselbeck, T.; Thielecke, H.; Köchlin, J.; Cho, S.; Weinschenk, I.; Metz, J.; Borggrefe, M.; Haase, K.K.

:

Basic research in cardiology 100 (2005), Nr.5, S.446-452
ISSN: 0300-8428
Englisch
Zeitschriftenaufsatz
Fraunhofer IBMT ()

Abstract
Newer techniques are required to identify atherosclerotic lesions that are prone to rupture. Electric impedance spectroscopy (EIS) can characterize biological tissues by measuring the electrical impedance over a frequency range. We tested a newly designed intravascular impedance catheter (IC) by measuring the impedance of different stages of atherosclerosis induced in an animal rabbit model. Six female New Zealand White rabbits were fed for 17 weeks with a 5% cholesterol-enriched diet to induce early forms of atherosclerotic plaques. All aortas were prepared from the aortic arch to the renal arteries and segments of 5-10 mm were marked by ink spots. A balloon catheter system with an integrated polyimide-based microelectrode structure was introduced into the aorta and the impedance was measured at each spot by using an impedance analyzer. The impedance was measured at frequencies of 1 kHz and 10 kHz and compared with the corresponding histomorphometric data of each aortic segment. Forty-four aortic segments without plaques and 48 segments with evolving atherosclerotic lesions could be exactly matched by the histomorphometric analysis. In normal aortic segments (P0) the change of the magnitude of impedance at 1 kHz and at 10 kHz (

: http://publica.fraunhofer.de/dokumente/N-45689.html