Fraunhofer-Gesellschaft

Publica

Hier finden Sie wissenschaftliche Publikationen aus den Fraunhofer-Instituten.

Cantilever micro balance as nanoparticle exposure monitor

 
: Wasisto, H. S.; Waag, A.; Merzsch, S.; Uhde, E.; Peiner, E.

Kızıl, Hüseyin:
MME 2014, 25th Micromechanics and Microsystems Europe Workshop. CD-ROM : August 31 - September 3, 2014, Istanbul, Turkey
Istanbul, 2014
4 S.
Micromechanics and Microsystems Europe Workshop (MME) <25, 2014, Istanbul>
Englisch
Konferenzbeitrag
Fraunhofer WKI ()
engineered airborne nanoparticles; cantilever resonator; electrophoretic sampler; personal monitor; direct read out; pocket size system

Abstract
An electrophoretic cantilever-based nanoparticle (NP) sensor was described for personal monitoring of occupational exposure at workplace. Online measurements performed under defined NP concentrations in a climate chamber confirmed the feasibility of the measurement principle. Cyclic switching between NP sampling and frequency tracking revealed settling times of 5 min and 1.5 min,respectively, until stable conditions were reached. Using an improved design of the sampling head a fully integrated pocket-size system offers a reduced response time of 20 s.