Fraunhofer-Gesellschaft

Publica

Hier finden Sie wissenschaftliche Publikationen aus den Fraunhofer-Instituten.

Calibration methods for high-g accelerometers

 
: Külls, Robert; Dessonet, Henri; Bohland, Christian; Nau, Siegfried; Thoma, Klaus

:

Institute of Electrical and Electronics Engineers -IEEE-:
Transducers 2013, 17th International Conference on Solid-State Sensors, Actuators and Microsystems : Eurosensors XXVII; Barcelona, Catalonia (Spain); 16 - 20 June 2013
Piscataway, NJ: IEEE, 2013
ISBN: 978-1-4673-5981-8
ISBN: 978-1-4673-5983-2
S.562-565
International Conference on Solid-State Sensors, Actuators and Microsystems (Transducers) <17, 2013, Barcelona>
European Conference on Solid-State Transducers (Eurosensors) <27, 2013, Barcelona>
Englisch
Konferenzbeitrag
Fraunhofer EMI ()

Abstract
High-g accelerometers are calibrated by dynamically loading them with sub-millisecond transient mechanical pulses and comparing their output with a reference measurement system. Usually, the determination of sensitivity is conducted by a peak value comparison. In this paper we present a calibration method that evaluates and compares all points of the measured pulse, thus leading to more calibration data per experiment. The method is applied to different reference measurement methods used during Hopkinson bar trials in the range of 20,000 g.