Fraunhofer-Gesellschaft

Publica

Hier finden Sie wissenschaftliche Publikationen aus den Fraunhofer-Instituten.

Biochip readout system for point-of-care applications

 
: Brandenburg, A.; Curdt, F.; Sulz, G.; Ebling, F.; Nestler, J.; Wunderlich, K.; Michel, D.

:

McDonagh, C.:
9th European Conference on Optical Chemical Sensors and Biosensors 2008. Proceedings : Dublin, Ireland, 30 March - 2 April 2008
Amsterdam: Elsevier, 2009 (Sensors and actuators. B, Chemical 139.2009, Nr.1)
ISSN: 0925-4005
pp.245-251
European Conference on Optical Chemical Sensors and Biosensors (Europt(r)ode) <9, 2008, Dublin>
English
Conference Paper, Journal Article
Fraunhofer IPM ()
Fraunhofer IZM ()
Fraunhofer ENAS ()
Fraunhofer IBMT ()

Abstract
A new low cost biochip readout system based on total internal reflection fluorescence (TIRF) including microfluidics is presented. The system is suitable for point-of-care diagnostics employing disposable substrates made of polymer foils producible at very low costs. Results for the reader sensitivity and dynamic range in comparison to state of the art readers are presented. The sensitivity is comparable to laser scanner systems developed for lab applications, although the newly developed TIRF system is built up much simpler and low cost substrates are used. The dynamic range and the on chip homogeneity were investigated. As an example for future POC applications, an immunoassay for the inflammation parameter C-reactive protein (CRP) is investigated. The sensitivity of these assays is in the range of 1 ng/ml, being sufficient for many parameters of clinical relevance.

: http://publica.fraunhofer.de/documents/N-96123.html