Comparative testing of the MCA-527 and MCA-166 mini multi channel analysers

 
: Köble, Theo; Risse, Monika; Schumann, Olaf; Jöster, Michael; Heinrichs, Matthias; Dürr, Martin; Ossowski, Stefan

:
Fulltext urn:nbn:de:0011-n-4815399 (682 KByte PDF)
MD5 Fingerprint: a46f2da2df1141120d376763531fcc2d
Created on: 30.1.2018

presentation urn:nbn:de:0011-n-481539-16 (498 KByte PDF)
MD5 Fingerprint: e6bc0e6f51766586e796878d0b823060
Created on: 30.1.2018


Sevini, Filippo (ed.) ; European Commission, Joint Research Centre -JRC-:
39th ESARDA Symposium on Safeguards and Nuclear Non-Proliferation 2017 : Düsseldorf, Germany, 16-18 May 2017
Luxembourg: Publications Office of the European Union, 2017
ISBN: 978-92-79-73861-6
pp.118-125
Symposium on Safeguards and Nuclear Non-Proliferation <39, 2017, Düsseldorf>
English
Conference Paper, Electronic Publication
Fraunhofer INT ()
gamma measurement; multichannel analyzer; comparative study

Abstract
This paper presents a comparative study of the two multichannel analysers MCA-166 and its successor the MCA-527 from GBS electronic with the focus of the operation at high counting rates. The MCA-166 is widely used by the IAEA and Euratom for safeguards inspections in the field. The performed tests included the influence of peak shaping parameters, count rate and temperature on the operation of the two multichannel analysers.