Options
2017
Journal Article
Titel
Germanium "hexa" detector. Production and testing
Abstract
Here we present new result on the testing of a Germanium sensor for X-ray radiation. The system is made of 3 × 2 Medipix3RX chips, bump-bonded to a monolithic sensor, and is called "hexa". Its dimensions are 45 × 30 mm2 and the sensor thickness was 1.5 mm. The total number of the pixels is 393216 in the matrix 768 × 512 with pixel pitch 55 m m. Medipix3RX read-out chip provides photon counting read-out with single photon sensitivity. The sensor is cooled to −126°C and noise levels together with flat field response are measured. For −200 V polarization bias, leakage current was 4.4 mA (3.2 m A/mm2). Due to higher leakage around 2.5% of all pixels stay non-responsive. More than 99% of all pixels are bump bonded correctly. In this paper we present the experimental set-up, threshold equalization procedure, image acquisition and the technique for bump bond quality estimate.