A cw-terahertz gas analysis system with ppm detection limits

 
: Hepp, C.; Lüttjohann, S.; Roggenbuck, A.; Deninger, A.; Nellen, S.; Göbel, T.; Jörger, M.; Harig, R.

:

Institute of Electrical and Electronics Engineers -IEEE-:
41st International Conference on Infrared, Millimeter, and Terahertz waves, IRMMW-THz 2016 : 25-30 September 2016, Copenhagen, Denmark
Piscataway, NJ: IEEE, 2016
ISBN: 978-1-4673-8485-8 (Electronic)
ISBN: 978-1-4673-8486-5 (Print on Demand)
pp.1263-1264
International Conference on Infrared, Millimeter, and Terahertz Waves (IRMMW-THz) <41, 2016, Copenhagen>
English
Conference Paper
Fraunhofer HHI ()

Abstract
We present a demonstration system for gas analysis based on cw-terahertz absorption spectroscopy. The setup combines tunable 1.5 μm distributed-feedback (DFB) lasers, InGaAs photomixers and a multipass absorption cell into a high-resolution spectrometer. At gas pressures of 1 mbar, we achieved detection limits in the range of approx. 20 ppm for ammonia (NH3) and 100 ppm for sulfur dioxide (SO2). Absorption profiles with halfwidths of 7 MHz (2×10-4 cm-1) were measured.