Development of an artifact-free aneurysm clip

Concept and approach for the development of an aneurysm clip entirely made from fiber-reinforced plastic materials
 
: Brack, Alexander; Senger, Sebastian; Fischer, Gerrit; Janssen, Henning; Oertel, Joachim; Brecher, Christian

:
Fulltext urn:nbn:de:0011-n-4289766 (672 KByte PDF)
MD5 Fingerprint: 9d14e53fb1c6a766e49db5d74a67b243
(CC) by-nc-nd
Created on: 14.1.2017


Current directions in biomedical engineering 2 (2016), No.1, pp.543-546
ISSN: 2364-5504
English
Journal Article, Electronic Publication
Fraunhofer IPT ()
Faserverbundkunststoff; Medizintechnik; Produktentwicklung; Produktionstechnik

Abstract
For the treatment of intracranial aneurysms with aneurysm clips, usually a follow-up inspection in MRI is required. To avoid any artifacts, which can make a proper diagnosis difficult, a new approach for the manufacturing of an aneurysm clip entirely made from fiber-reinforced plastics has been developed. In this paper the concept for the design of the clip, the development of a new manufacturing technology for the fiber-reinforced components as well as first results from the examination of the components in phantom MRI testing is shown.