Standard free thickness determination of thin TEM samples via backscatter electron image correlation

 
: Salzer, R.; Graff, A.; Simon, M.; Altmann, F.

Kothleitner, G.:
MC 2009. Proceedings. Vol. 1: Instrumentation and methodology : Microscopy Conference, Graz, Austria, 30 August - 4 September 2009. First joint meeting of Dreiländertagung & Multinational Congress on Microscopy
Graz: Verlag der TU Graz, 2009
ISBN: 978-3-85125-062-6
pp.237-240
Microscopy Conference (MC) <2009, Graz>
English
Conference Paper
Fraunhofer IWM ()

Abstract
Due to the ongoing progress in nanotechnology, high resolution TEM investigations become more and more important in structural and chemical material analyses. For such investigations the thickness of the TEM sample should be as thin as possible. The thickness of TEM samples prepared by focused ion beam (FIB) is about 50 to 100 nm. For light elements like silicon this might be sufficient for most of the analytical work. Due to the Z-dependence of the electron scattering it is also necessary to achieve even smaller thicknesses to perform TEM analyses at materials with high mass densities or high atomic numbers like Au or Pt. The difficulties to prepare plane-parallel lamellae with thicknesses of 10 to 50 nm by FIB are known. Common methods especially the FIB preparation do not have the abi lity to control the thickness precisely during the preparation process. Therefore, it is difficult or nearly impossible to stop the FIB sputtering process at the right target thickness. The aim of the presented work was to use the contrast formation of backscattered electrons (BSE) at thin samples by SEM during FIB preparation to determine the thickness of the observed TEM sample. The images produced by backscattered electrons have the advantage that they are not influenced by any FIB generated electrons. Thus, BSE can be used to monitor the thickness changes whilst FIB milling of TEM samples. The contrast in a backscattered electron image of a thin sample depends among other things on the maximal depth from which electrons could be backscattered. If the thickness of a sample is smaller than this depth a part of the primary electrons, which would be backscattered in bulk samples, will traverse the sample and could not be detected as BSEs. So the amount of BSEs will decrease (see thin edge of the wedge in Fig. 1). This thickness dependent contrast change is also a function of the material of the sample and the energy of the incident electrons. To describe this BSE contrast formation we derive a global normalized backscatter electron transmission function (GNTF). This function depends on primary electron energy, thickness and material parameters (element and density) of the lamella. It is based on a modified regularized incomplete Gamma-function adapted to Monte-Carlo simulations of electron trajectories in solids (CASINO) [1] (Fig. 1 b). The fact that BSE have a depth dependent energy distribution and the used BSE detector have an enery dependent amplification is considered by a correction function which is based on a continuous slowing down approximation. The GNTF and the correction function are used to calculate for each pixel of a BSE-image of a thin sample a thickness value and display all of them in a color-coded thickness map. For this the transmission function is scaled with the maximum (bulk sample area) and the minimum (background without sample) grey values (Fig. 1 a). To verify the calculated normalized BSE transmission gradients, wedges of different materials were prepared by FIB milling. The thickness determined by analyzing the grey values was compared to values measured at cross sections at certain markers (Fig. 1, 2).

: http://publica.fraunhofer.de/documents/N-104803.html